Pig ACE(Angiotensin I Converting Enzyme) ELISA Kit – ATO5772
ATO5772
Size:Price:
48 T$412.00
96 T$588.00

Overview

Product name: Pig ACE(Angiotensin I Converting Enzyme) ELISA Kit
Reactivity: Pig
Alternative Names: CD143; ACE1; DCP1; ACEI; ACE-I; Kininase II; Angiotensin-Converting Enzyme; Peptidyl-Dipeptidase A; Dipeptidyl Carboxypeptidase 1; Angiotensin-converting enzyme, soluble form
Assay Type: Sandwich
Sensitivity: 11.2 pg/mL
Standard: 2000 pg/mL
Detection Range: 31.25-2000 pg/mL
Sample Type: Serum, plasma and other biological fluids.
Assay Length: 3.5h
Research Area: Enzyme & Kinase;Cardiovascular biology;
Uniprot ID: F1RRW5
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Pig ACE. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Pig ACE. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Pig ACE, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm 10nm. The concentration of Pig ACE in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
2000.00 2.201 2.106
1000.00 1.579 1.484
500.00 1.121 1.026
250.00 0.933 0.838
125.00 0.572 0.477
62.50 0.389 0.294
31.25 0.215 0.120
0.00 0.095 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant ACE and the recovery rates were calculated by comparing the measured value to the expected amount of ACE in samples.
Matrix Recovery range Average
serum(n=5) 95-107% 101%
EDTA plasma(n=5) 80-95% 87%
Heparin plasma(n=5) 88-103% 95%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ACE and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-103% 90-99% 87-95% 89-97%
EDTA plasma(n=5) 95-105% 82-95% 81-97% 93-105%
Heparin plasma(n=5) 85-97% 90-99% 82-101% 93-105%